1. An automatic, electromagnetic diver balance is described.

  2. The working principle of this instrument is that a small plastic cylinder, the diver, with density less than one is kept floating in an aqueous medium by a magnetic force acting on a permanent magnet enclosed in the diver. In the floating position the diver cuts off a light beam which, through an electronic system, controls the magnetic force acting upon the diver. The latter will move down until its buoyancy is compensated by this force.

  3. With this balance one may register the submerged—or reduced—weight of any object placed on the floating diver.

  4. The apparatus has hitherto found application in water-permeation studies (D2O-H2O exchange) and in determinations of the cortical tension in the frog’s egg. Potentially the instrument may be used for various other purposes, notably for manometric determinations.

    Miss Rita Gustafson, Mrs Margareta Jaksch, and Mrs Inger Janson have done all the practical work associated with testing out the method, their excellent assistance is gratefully acknowledged. The work has been supported by grants from Natur-vetenskapliga Forskningsrådet.

The diver balance constructed by Zeuthen (1948) was based on the Cartesian principle, i.e. that the specific gravity of a floating body containing an air bubble may be changed by variation of the pressure above the surface of the floatation medium. This instrument was a further development of the Cartesian diver, the microrespiro-meter developed by Linderstram-Lang (1943) and Holter (1943).

The diver balance is used for weighing objects submerged in the floatation medium, and it can therefore determine only the submerged or—to use an expression coined by Linderstrom-Lang & Holter (1940)—the reduced weight (RIP).

RW determinations have been of good service in the solution of various biological problems, e.g. for reference to the size of biological objects (cf. Avery & Linderstram-Lang, 1941 ; Holter & Løvtrup, 1949), and in studies of the utilization of endogenous energy sources in starving organisms (cf. Holter & Zeuthen, 1948; Lavtrup, 1959). The use of the diver balance for studies of water permeability, using D2O as a tracer, was introduced by Pigon & Zeuthen (1951) and Løvtrup & Pigon (1951).

The diver balance has also been used in studies on eggs of marine animals. Thus Zeuthen (1951) studied the changes in RW during the early stages of sea-urchin development. Rhythmical changes, consonant with the mitotic rhythm, were observed, but these changes were superimposed upon a very extensive change in RW, apparently not associated with changes in the living material. Some years ago experiments were initiated to determine the rate of water exchange in sea-urchin embryos. During this work Zeuthen’s observation was confirmed; it could actually be established that if the equilibrium of an unloaded balance was upset by a large change of presssure in the floatation vessel, then the diver balance would float at a pressure widely different from the original one. A gradual change occurred, however, and after about half an hour the original level was reached and no further changes were observed.

Subsequent control experiments have shown that this phenomenon occurs even in fresh water, but the absolute change is slight, and the original value is reached in a few minutes. Consequently the measurement of water exchange in fresh water is not seriously interfered with.

It is evident that, whatever the reason of the observed RIP-change—various suggestions have been made, but no definite explanation seems agreed as yet—the Cartesian diver balance can never be used for water-exchange studies in salt water. A diver balance, working on some completely different principle, is required for this purpose. Since the difficulties encountered. with the manometric diver balance probably are concerned with the exchange between the enclosed air bubble and the surrounding medium, the electromagnetic balance devised by Brzin, Kovič and Oman (1964) may meet this demand.

In permeability studies the time course of the RW changes must be recorded in order to establish the rate constant for the water-exchange reaction. Simultaneous adjustment and observation of the floating position of the diver, reading of equilibrium pressure and time are such complicated operations that the participation of two technicians is required, at least during the early phase of the process when the rate of change is very fast.

It was from the outset obvious to us that if any substantial improvement of the diver balance technique were to be achieved, it would be necessary to do away with this tedious work.

This requires not only that the readings be automatically recorded—this can be done with the electromagnetic balance mentioned above—but also that the control of the floating level be automatic. The electromagnetic diver balance presented here fulfils both demands.

Construction of the balance

Principle of operation

The electromagnetic diver balance is based on the principle that a balance pan— the diver—with specific gravity less than the surrounding medium is prevented from rising to the surface by an electromagnetic force acting on a small magnet contained in the diver. The adjustment of the equilibrium position is achieved by means of a servo-loop, comprising an optical, an electronic and a magnetic component (Fig. 1).

Fig. 1.

Diagram of the operating principle of the electromagnetic diver balance.

Fig. 1.

Diagram of the operating principle of the electromagnetic diver balance.

The light source, a lens bulb (Osram 3644), sends an almost parallel beam of light through a cuvette filled with an appropriate floatation medium, and through a narrow slit in the photocell housing to the photocell (Siemens photodiode APY 11 I). The latter controls by means of an amplifier the current passing through an electromagnet placed under the cuvette. When the diver is pulled down by the current, its lower end will cut off the light to the photocell. Since the current in the coil is proportional to the light reaching the photocell the force will diminish as the diver moves down.

When the electromagnetic force exactly balances the upward force of the diver the latter has reached its equilibrium position, and no further changes occur.

Upon loading, the equilibrium force will decrease and the diver will accordingly sink until a new equilibrium is reached.

Arrangement of optical and magnetic components

The weighing system is very sensitive to temperature changes, and the cuvette, lamp housing, photocell housing and coil are therefore immersed in a thermostat-regulated water bath with efficient circulation. In very sensitive work we have found that an inner bath, of Plexiglass, may be necessary. The water in this bath is not heated, but it is stirred by a circulation pump.

The mechanical construction of the various units is shown in Fig. 2.

Fig. 2.

Cross-section of the submerged part of the electromagnetic diver balance. A, cuvette; B, photocell housing; C, photocell diode; D, o-rings; E, Plexiglass windows; F, lamp housing ; G, lamp ; H, Plexiglass plate ; I, electromagnet ; J, centrehole for iron core ; K, cuvette support.

Fig. 2.

Cross-section of the submerged part of the electromagnetic diver balance. A, cuvette; B, photocell housing; C, photocell diode; D, o-rings; E, Plexiglass windows; F, lamp housing ; G, lamp ; H, Plexiglass plate ; I, electromagnet ; J, centrehole for iron core ; K, cuvette support.

The lamp housing and the photocell housing are made of nickel-plated brass. The small windows are made of Plexiglass, provided with grooves for o-rings ; great care must be devoted to the finish of the surfaces in order to keep down the losses of light. The slit (0·5-1 × 2-3 mm.) in the window of the photocell housing reduces the width of the light beam, and thus permits the use of thinner divers.

The electromagnetic coil consists of 0·55 mm. copper wire, 2000 turns wound on a Plexiglass spool with a central opening of 10 mm. A wide Plexiglass tube covers the spool ; the tightening is achieved with o-rings. If required an iron core may be placed in the central hole.

All the submerged components are attached to a 12 mm. Plexiglass plate, including the cuvette holder. Allowance is made for adjustment to varying cuvette sizes.

Electronic circuits

The servo amplifier (Fig. 3) is designed as a transistorized a.c.-amplifier with an output of 10 W. The photodiode is fed by pulses from a pulse generator (1000 cyc./sec.). The amplitude of this signal is modulated by the light reaching the photocell, and transferred to the a.c.-amplifier which, through a transformer and a rectifier, is coupled to the coil. The latter is thus being fed with a d.c.-potential. The input signal may be regulated by a potentiometer, by which the floating level may be adjusted. The signal from the photodiode is also fed to a circuit containing a meter on which may be read the relative height of the diver in the light beam, i.e. the floating level.

Fig. 3.

Block diagram of the electronic circuits in the electromagnetic diver balance.

Fig. 3.

Block diagram of the electronic circuits in the electromagnetic diver balance.

The measuring circuit contains two 10-turn helical potentiometers, by which the sensitivity and the zero-position on the recorder may be adjusted. A 10 mV potentio-meter recorder—multichannel if desired—is used for recording the changes in the equilibrium current.

In cases where no continuous changes are to be recorded, a simple zero instrument may replace the recorder. Reading of the equilibrium current is then made on the zero-position potentiometer.

The diver

The divers are made of polypropylene, which may be obtained as wires and sticks of varying width. Most of the divers prepared so far have been made of 3-4 mm. wire. Suitable pieces are cut off and placed in a lathe. In order to get the lower end as plane as possible, 1-2 mm. is cut off from the free end with a sharp knife, while the lathe is rotating. Subsequently a hole is drilled in the piece, 0·1-0·2 mm. wider and 1 mm. longer than the magnetic material to be inserted.

The magnets are made out of 0·5 mm. Koerzit T*; except for special purposes pieces of 1-3 mm. were employed. The pieces are hardened in an oven at 530-570° C. ; after cooling they are put into the hole in the plastic and the opening is fused with a hot wire. Subsequently the material is magnetized by placing the diver in a magnetic field, the strength of which is adjusted according to the desired sensitivity of the diver. The lower end of the diver is then cleaned with a knife, and ground with various grades of oil-resistant carborundum paper, to make the surface as smooth as possible.

The loading capacity of the diver (= — RWd, where RWd is the reduced weight of the diver, cf. below) will be determined by the amount of polypropylene. This cha-racteristic may be adjusted by cutting off slices from the upper end. An approximate estimate of the capacity can be made by putting the diver in a beaker with water and loading it with known weights.

As we shall discuss later, the sensitivity of the balance varies inversely with the ‘magnetic quantity ‘in the diver ; in order to make a sensitive diver the magnet must therefore be very small. Under these circumstances the amount of plastic required to give a reasonable RWdmay be quite low. In order to bring the diver to a convenient size a small piece of platinum wire may be introduced below the magnet before the hole is closed.

It has been observed that the divers change their RWd in water, especially during the initial period. This change is rather slow, and may easily be corrected for.

A loaded floating diver is shown in Fig. 4.

Fig. 4.

Photograph showing part of the diver balance set-up. In the cuvette a diver is floating, loaded with a frog’s egg.

Fig. 4.

Photograph showing part of the diver balance set-up. In the cuvette a diver is floating, loaded with a frog’s egg.

Weighing theory

Loading capacity J

The electromagnetic diver balance is used to determine the reduced weight
where V and ρ represent volume and specific gravity, and the suffixes o and m represent object and medium, respectively.

In the Cartesian diver balance the RW of the diver proper is a measure of the sensitivity of the balance (cf. Løvtrup, 1951). The loading capacity depends on details of the construction of the apparatus, viz. the space available for expansion of the air bubble, and the range of the manometer.

In the electromagnetic diver balance the opposite relation holds ; here the numerical value of the reduced weight of the diver, RWd, represents the maximum load, the loading capacity Lmax. The sensitivity may, on the contrary, be changed by adjustment of the amplifying system.

In order to use a diver it must be possible for the magnet coil to balance the reduced weight of the empty diver. If by F we designate the electromagnetic force acting on the diver, we have F = KoiM., where Ko is a constant incorporating features of the design of the apparatus, i the current in the coil, and M the compound magnetic quantity of the diver magnet. This latter may be further resolved since
where Mo is the magnetic quantity in the poles of the magnet, m the length of it, and I the distance from the coil to the lower end of the magnet. It is a design condition that
The right-hand inequality states that the diver must be able to float. It is seen that the maximum value of RWd under otherwise constant conditions will be determined by the maximum current in the coil. The range may be changed by variation of some of the factors responsible for the strength of the electromagnetic force, all of which are incorporated in M and Ko, e.g. the distance between coil and magnet, presence or absence of an iron core in the electromagnet, etc.
From the inequality (3) it is seen that the ratio between the loading capacity and the magnetic quantity in the diver cannot exceed a certain value
Since it is advantageous to work with low currents when possible, it may be advisable to use divers with a capacity not too much above the RW of the object. This will require a set of divers covering different ranges. It is also possible to start with a diver with high capacity; this may then be reduced to different levels by placing platinum wires of varying RW on the diver.

In some cases it may be expedient to work with a diver having so large capacity that it cannot be equilibrated when empty. This situation arises if in a heavy object relatively small changes are to be measured. The recording can then be made only after the object has been placed on the diver. As we shall see, the calibration is linear as long as the displacement of the diver is small compared with I, so that this procedure will consequently entail no changes as concerns the conversion of the readings to absolute units.

It will be realized from the definition of R W that the capacity is a function of both volume and specific gravity of the diver. There is thus no direct proportionality between the size of the diver and the capacity.

It is not easy to determine these parameters with sufficient accuracy for calculating the RWd. Approximate determination of the capacity may be achieved by loading with gradually increasing loads until the diver sinks. A somewhat more accurate method will be described below. However, since this parameter is of no importance for the calibration of the balance, an exact determination of the capacity is in general of no particular interest.

Sensitivity and stability

The sensitivity of the electromagnetic balance, i.e. the recorder deflexion per unit load, may be changed by the setting of the sensitivity knob on the amplifier. However, by this expedient the background noise is changed correspondingly; at high sensitivities this factor thus becomes of increasing importance. It may therefore be worth the while to investigate whether it is possible to influence the sensitivity through details in the construction of the balance or in the design of the diver. The absolute sensitivity, i.e. the change in current for a given load, may be derived from the equation representing the equilibrium condition of a floating, loaded diver.
where RWx represents the reduced weight of a load placed on the balance. For the sensitivity we get
This expression contains a negative sign since the current increases as the load decreases. The sensitivity is seen to depend on Ko and M, factors that are constant for a particular set-up. The smaller the values of these factors, the more sensitive is the balance. The sensitivity is seen to be decreased by placing an iron core in the magnetic coil, since this will increase the value of Ko, which is proportional to the ratio between the electromagnetic force and the current in the coil. From equation (2) it is seen that M, the compound magnetic quantity in the diver, will decrease, and the sensitivity will consequently increase, when the magnet in the diver is weak and short, and when the distance to the electromagnet is long.

Thus, in cases where very high sensitivity is required, it is necessary to use a diver with a weak magnet, but adjustment of the other factors mentioned above may also help to reach the required sensitivity. All these factors that tend to improve the sensitivity will at the same time reduce the maximum value of RWd, but it is important to note that the loading capacity of the diver, — RWd, has no influence on the sensitivity.

The equilibrium position, and thus indirectly the sensitivity of the floating diver, may be influenced by a variety of factors ; one of the most important is probably the specific gravity of the medium, not the least because thjs factor is so sensitive to changes in temperature. In order to investigate whether it is possible in the design of the diver to take precautions to increase the stability, we rewrite equation (5) as follows :
For the stability with respect to changes in the specific gravity of the medium we get
Thus, the more we improve the sensitivity by decreasing Ko and M, the more unstable the diver becomes. If these values are fixed, we may improve the stability by decreasing Vd, the volume of the diver. The design of the diver must thus be a compromise between the requirements for sensitivity and stability.

Correlation between floating level and current

The electromagnetic force acting on the empty floating diver is equal to — RWd. As the diver is loaded the force required to keep the system floating diminishes. The diver sinks in the container until the current through the magnet coil suffices to balance the RWof balance + load. When the load is equal to — RWd this force, and thus the current, is zero. This situation obtains when the diver just cuts off the whole light beam. This relation is shown in Fig. 5. Since amplification is linear, the ratio between diver displacement and current change is also linear. The slope of the line may be changed by the floating-level adjustment knob, controlling the amplification. The floating position of the empty diver will change correspondingly, as shown in the figure. It will be realized that the absolute value of Δh, the change in position resulting from loading the diver, will be changed proportionally by this expedient.

Fig. 5.

Correlation between load, current in the electromagnetic coil and the floating level of the diver. The two sloping lines correspond to different settings of the floating-level adjustment knob.

Fig. 5.

Correlation between load, current in the electromagnetic coil and the floating level of the diver. The two sloping lines correspond to different settings of the floating-level adjustment knob.

The distance l between the magnet and the coil will be changed by the amount Δh when the floating position is changed. This fact will cause a deviation from linearity in the calibration curve. The ratio AA/Z is very small, however, and may be further reduced by increasing the amplification. In our work with the diver balance we have adjusted the magnification so that no significant deviation from linearity obtains (cf. Fig. 6).

Fig. 6.

Calibration curve.

Fig. 6.

Calibration curve.

If the amplification is increased too much the effect of disturbing influences of various kinds, e.g. convection currents, particles in the light path, etc., will also be magnified, since a small displacement of the diver will entail a large change in current.

At high magnifications the risk also increases that the diver may begin to oscillate. This may be prevented by raising the viscosity of the medium and enlarging the diameter of the diver, or electronically by a negative feed-back system. In the work carried out with the diver balance so far, we have had no trouble with self-oscillations, and none of the mentioned expedients have therefore been introduced, but it may be necessary to consider these possibilities under different experimental conditions.

Under any circumstances it will be realized that in practical work with the diver balance the degree of magnification must strike a practical compromise between the demands for linearity and stability.

Calibration

Neglecting the displacement Δh we have for the calibration (equation 6)
A linear relationship thus obtains between current and load. This point has been tested experimentally by loading a diver with known weights made out of platinum wire (d = 100 μ). The results shown in Fig. 6, demonstrate that within the limits of error the linearity was confirmed.
With this relation established it is possible to determine RWd in a way that is potentially quite accurate. The floating-level adjustment knob is connected with a meter showing the percentage of the light reaching the photocell. If the amount cut off by the unloaded diver is x1, and after loading with is x2, it appears from Fig. 5, that we may write

Weighing procedure

Before loading, the empty diver is made to float, and the zero-position knob is adjusted to move the recorder to the base-line. The reading on the zero position potentiometer P0 will be proportional to the current in the coil
After loading a deflexion D of the recorder occurs. If the sensitivity adjustment, S, is appropriate, the deflexion remains within the recorder range. We thus have
In some cases it may be advantageous to work with a sensitivity too high to permit the deflexion to remain within the recorder range. This is particularly desirable when RW changes are to be recorded that are only a fraction of the total load. Under these conditions a second adjustment of the zero position must be made, P1. We thus obtain
Combining (10) and (12) we get
The values of K3 and Kt may be determined by loading the diver balance with known weights. K3 is then obtained from the change in the zero-position knob, required to keep the recorder at the base-line, whereas Kt is obtained from the recorder deflexion when the zero position is unchanged.

Application

The electromagnetic diver balance may be used for the same purposes as the Cartesian diver, for weighings in general, including metabolic studies on starving organisms, and for water-exchange studies (cf. fig. 7). As far as the last application is concerned both diffusion and permeability coefficients may be determined (cf. Løvtrup, 1963). Obviously these determinations are not necessarily limited to work with biological objects, or to water exchange. It may just as well be possible to follow the diffusion of organic solvents in various plastic objects, etc.

Fig. 7.

The H2O-D2O exchange in a frog’s egg, recorded by means of the electromagnetic diverbalance. Abscissa: time, increasing from right to left. Ordinate: potentiometer deflexion.

Fig. 7.

The H2O-D2O exchange in a frog’s egg, recorded by means of the electromagnetic diverbalance. Abscissa: time, increasing from right to left. Ordinate: potentiometer deflexion.

The electromagnetic diver balance may also be used for manometric studies, thus substituting for the Cartesian diver. If an object is placed in a drop in a capillary, surrounded by air bubbles, then any consumption or production of gas will change the volume of the bubbles, and consequently the R W of the capillary. If the latter is placed on the diver, it will thus be possible to follow the gasometric reaction. By appropriate modification and enlargement of the present model it should be possible to construct an apparatus that might to all intents and purposes function as an automatically recording Warburg apparatus.

The electromagnetic diver balance may also be used to determine various kinds of forces and tensions. In the method of Cole (1932) the mechanical properties of the cortex in the sea-urchin egg are determined from the deformation of the egg arising after application of a known compressive force (cf. also Hiramoto (1963) and Yoneda (1964)). In a study of the tension in amphibian eggs the diver balance has successfully replaced the torsion-beam balance employed by Hiramoto (Bemtsson, unpublished results).

Avery
,
G. S.
Jr
. &
Linderstrom-Lang
,
K.
(
1941
).
Peptidase activity in the Avena coleoptile, phyto-hormone teat object
.
C.R. Lab. Carlsberg, (Ser. chim.)
,
23
,
219
34
.
Brzin
,
M.
,
Kovic
,
M.
&
Oman
,
S.
(
1964
).
The magnetic diver balance
.
C.R. Lab. Carlsberg
,
34
,
407
26
.
Cole
,
K. S.
(
1932
).
Surface forces of the Arbacia egg
.
J. Cell. Comp. Physiol
.
1
,
1
9
.
Hiramoto
,
Y.
(
1963
).
Mechanical properties of sea urchin eggs. 1. Surface force and elastic modulus of the cell membrane
.
Exp. Cell. Res
.
32
,
59
75
.
Holter
,
H.
(
1943
).
Technique of the Cartesian diver
.
C.R. Lab. Carlsberg (Sér. chim.)
,
24
,
399
478
.
Holter
,
H.
&
Lovtrup
,
S.
(
1949
).
Proteolytic enzymes in Chaos chaos
.
C.R. Lab. Carlsberg (Sér chim.)
,
27
,
27
62
.
Holter
,
H.
&
Zeuthen
,
E.
(
1948
).
Metabolism and reduced weight in starving Chaos chaos
.
C.R. Lab. Carlsberg (Sér chim.)
,
26
,
277
96
.
Linderstrom-Lang
,
K.
(
1943
).
On the theory of the Cartesian diver microrespirometer
.
C.R. Lab. Carlsberg (Sér chim.)
,
24
,
333
98
.
Linderstrom-Lang
,
K.
&
Holter
,
H.
(
1940
).
In Bamann and Myrback
,
Die Methoden der Ferment-forschung
.
Leipzig
.
Lovtrup
,
S.
(
1951
).
Observations on the Cartesian diver balance technique
.
C.R. Lab. Carlsberg. (Sér. chim.)
,
27
,
125
36
.
Lovtrup
,
S.
(
1959
).
On the causes of variations in the respiratory curves during amphibian embryo-genesis
.
J. Exp. Zool
.
140
,
231
46
.
Lovtrup
,
S.
(
1963
).
On the rate of water exchange across the surface of animal cells
.
J. Theoret. Biol
.
5
,
341
59
.
Lovtrup
,
S.
&
Picon
,
A.
(
1951
).
Diffusion and active transport of water in the amoeba Chaos chaos L
.
C.R. Lab. Carlsberg (Sér. chim.)
,
28
,
1
36
.
Picon
,
A.
&
Zeuthen
,
E.
(
1951
).
Cartesian diver balance in permeability studies
.
Experientia
,
7
,
455
.
Yoneda
,
M.
(
1964
).
Tension at the surface of sea-urchin egg: A critical examination of Cole’s experiment
.
J. Exp. Biol
.
41
,
893
906
.
Zeuthen
,
E.
(
1948
).
A Cartesian diver balance weighing reduced weights (R.W.) with an accuracy of±o-oiy
.
C.R. Lab. Carlsberg (Sér. chim.)
,
26
,
243
66
.
Zeuthen
,
E.
(
1951
).
Segmentation, nuclear growth and cytoplasmic storage in eggs of echinoderms and amphibia
.
Pubbl. Stax. Zool. Napoli
23
(
Suppl
.),
47
69
.
*

Supplied by Fried. Krupp Widia-Fabrik, Hessen, Western Germany.