ABSTRACT
We explore a recent, innovative variation of closed-system respirometry for terrestrial organisms, whereby oxygen partial pressure (PO2) is repeatedly measured fluorometrically in a constant-volume chamber over multiple time points. We outline a protocol that aligns this technology with the broader literature on aerial respirometry, including the calculations required to accurately convert O2 depletion to metabolic rate (MR). We identify a series of assumptions, and sources of error associated with this technique, including thresholds where O2 depletion becomes limiting, that impart errors to the calculation and interpretation of MR. Using these adjusted calculations, we found that the resting MR of five species of angiosperm seeds ranged from 0.011 to 0.640 ml g−1 h−1, consistent with published seed MR values. This innovative methodology greatly expands the lower size limit of terrestrial organisms that can be measured, and offers the potential for measuring MR changes over time as a result of physiological processes of the organism.
INTRODUCTION
One of the basic processes of life is the oxidation of substrates, termed metabolism (Kleiber, 1961). Simple measures of O2 consumption to infer metabolic rate (MR) are difficult to compare across diverse species because patterns of ecology and evolution influence both MR (Isaac and Carbone, 2010; White and Kearney, 2014) and its method of measurement (Lighton, 2008; Vleck, 1987; Withers, 2001). Calculating MR has a long heritage from Lavoisier in the 1770s (Hulbert and Else, 2004; Lighton, 2008), and there are many methods for quantifying MR, but indirect calorimetry measuring O2 consumption and/or CO2 production has gained ascendancy owing to the ease of measuring gas exchange by living organisms (Kleiber, 1961; Lighton, 2008; Withers et al., 2016).
Two respirometry methodologies (closed and open system) are based on a simple rationale – the measurement of O2 uptake and/or CO2 excretion over a known time course. In a closed system, the organism is kept in a sealed container of known air volume and the change in the O2 (and/or CO2) composition of air is measured over a time interval (Lighton, 2008; Vleck, 1987). Typically, closed systems are constrained to one-off measurements, from the start to end of a trial, allowing only an average MR to be calculated. Closed systems can be optimised to measure small organisms, either by reducing chamber volume or by extending measurement time to maximise the amount of gas exchange. Conversely, open, flow-through systems allow continual, real-time measurement of MRs in changing states of metabolic activity (Withers, 2001). Measurement of small organisms in an open system, however, requires low flow rates to allow O2 depletion levels to remain within detectable limits of the analyser, constraining open systems to a lower mass limit.
In both open and closed respirometry, the measurement of O2 partial pressure (PO2; kPa) has traditionally been made using paramagnetic (Hill, 1972; Nunn et al., 1964), fuel-cell (Torda and Grant, 1972) or polarographic gas analysers (Clark et al., 1953), whereas CO2 partial pressure (PCO2) measurements are determined using infra-red gas analysers (Fowler, 1949). More recently, a novel technique has emerged that measures PO2 by predictable and repeatable fluorometric quenching of metal organic dye located inside the respirometry chamber (e.g. Köster et al., 2008; Szela and Marsh, 2005; Bradford et al., 2013). Such fluorescent analysers have been used for several years to measure MRs of small, aquatic organisms (Alton et al., 2012; Bywater et al., 2014; Köster et al., 2008; Szela and Marsh, 2005) but less commonly for small, terrestrial or diving organisms (Bradford et al., 2013; Matthews and Seymour, 2010).
Typically, closed-system respirometry has been used to measure the PO2 change from the start to the end of a trial, by analysing an air sample from the chamber, and average MR is calculated (Vleck, 1987). A key advantage of fluorometric O2 measurement, however, is that the non-invasive nature of the in situ fluorescence dye allows for sequential, repeated measurement of the same organism in the sealed metabolic chamber (e.g. in aquatic systems; Alton et al., 2012; Szela and Marsh, 2005). The automation of such repeated measurements has recently been specifically developed for measuring O2 consumption in an aerial closed respirometry system of seeds (Bradford et al., 2013) and leaf tissue of Arabidopsis (Scafaro et al., 2017). Previous studies have used standardised O2 depletion as a proxy for MR of seeds from highly selectively bred cultivars; for example, in relation to the seed quality of tomato and lettuce varieties (Bello and Bradford, 2016; Bradford et al., 2013). Here, we refine the handling and processing of closed-system fluorometric data to calculate MR in traditional units (e.g. ml O2 h−1, or ml O2 g−1 h−1), which allows the quantitative comparative analysis of MR variation expected for diverse, wild organisms.
Making repeated measurements of PO2 and continuously calculating MR provides all the benefits normally ascribed to flow-through respirometry, including the ability to detect periods of altered metabolic states such as metabolic depression (Guppy and Withers, 1999). By aligning repeated-measurement fluorescence oxygen-sensing technology with broader respirometric theory and practice, we believe this represents a major advance in the measurement of MR for the great diversity of organisms too small for existing techniques. However, the location of the fluorescent sensor within a sealed aerial respirometry chamber is problematic, compared with conventional closed-system respirometry. Air sealed within the respirometric chamber is subject to PO2 changes not only by organismal metabolism but also by changes in air pressure within the sealed chamber (e.g. owing to ambient temperature, water vapour pressure and PCO2 changes). This substantially alters some of the assumptions and calculations of standard closed-system respirometry (Vleck, 1987).
- ATPS
ambient temperature and pressure, saturated with water vapour
- FeO2
subsequent (end) fractional O2 value
- FiO2
initial fractional O2 value
- FsO2
fractional standardised atmospheric O2
- MR
metabolic rate
- Pb
barometric pressure
- PCO2
CO2 partial pressure
- PO2
O2 partial pressure
- RER
respiratory exchange ratio
- RH
relative humidity
- RMR
resting metabolic rate
- STPD
dry standard temperature and pressure
- SWVP
saturation water vapour pressure
- Ta
chamber temperature
- V̇O2
volume of oxygen consumed (i.e. metabolic rate)
- vg
gas volume
Here, we have adapted several of the metabolic calculations of standard closed-system respirometry (Vleck, 1987) to align repeated-measurement fluorescence closed-system aerial respirometry with expectations of conventional closed system respirometry. Since repeated-measurement closed-system aerial respirometry has been exclusively used in the field of seed biology, we have used seeds as model species to apply these calculations for resting MR (RMR). Although the calculation of MR can provide insight into patterns and processes contributing to biological variation, miscalculation of MR can easily result from methodological and arithmetical errors (Lighton, 2008). We discuss these errors, and describe how they may be accounted for, eliminated or reduced in magnitude.
MATERIALS AND METHODS
Closed system respirometry
We used Q2 technology (ASTEC Global, The Netherlands) for repeated-measurement fluorometric respirometry of seeds in air-filled 1800 µl closed plastic vials (Bradford et al., 2013; Scafaro et al., 2017). The fluorometric O2 detector is suspended from a robotic arm above a series of vials laid out in a grid; the inner surface of the vial cap is impregnated with a fluorescent dye. Automated movement of the detector provides repeated measurement of internal PO2 for vials at user-determined intervals (Scafaro et al., 2017). The fluorometer is calibrated before each set of measurements using standards containing ambient air (designated as fractional standard atmospheric O2; i.e. FsO2=1.0) and O2-free air (FsO2=0.0) above a saturated sodium dithionite solution (Scafaro et al., 2017). All organismal O2 measurements are presented as standardised O2 fractions.
The calculations we report here account for chamber size. The Q2 system has standardised chamber types and sizes, ranging from 5500 µl chambers on a 24-well plate (384 chambers for 16 plates) to 500, 1800 and 2200 µl chambers on a 48-well plate (768 chambers for 16 plates), and 515 µl chambers on a 96-well plate (1536 chambers for 16 plates); we used 24- and 48-well plates. We make the general assumption that the respirometry chamber is sealed, impermeable to gases, and rigid (constant volume).
Calculation of MR
We outline here the calculation of absolute MR (V̇O2; ml O2 h−1) and highlight some assumptions that currently remain untested for repeated-measures closed-system fluorometric respirometry, modifying conventional closed respirometry calculations (Vleck, 1987). Automated repeated and long-duration measurements possible with fluorometric respirometry result in large volumes of raw, ‘noisy’ data for calculation. We have developed an R script that automates our calculations, constructed around the specific data file structure of the Q2 software, but these should be readily translatable to other data from other systems. The R script is available in the supplementary material, and is heavily annotated to provide guidance to its use (Script 1).
Respirometry chamber air volume
The initial chamber volume at dry standard temperature and pressure (STPD) conditions must be known, and corrected for the space occupied by the organism(s) in each chamber (Vleck, 1987).
where Ta is the initial chamber temperature (°C), Pb is the initial barometric pressure (kPa) and SWVP is the saturation water vapour pressure (kPa).
Changing air constituents and partial pressures
The measurement of PO2 by fluorometry occurs inside the sealed respirometry chamber, unlike conventional closed-system respirometry, where an air sample is removed from the chamber for analysis. Simplistically, the PO2 is dependent on barometric pressure inside the closed chamber at the time of measurement (Pb; kPa), but Pb inside the respirometry chamber is not known (or measurable) over repeated measurements. The calculation of PCO2 depends on the percent CO2 content of chamber air (which the Q2 does not measure), and chamber pressure. The partial pressure of N2 inside the respirometry chamber is not known, but we assume that the molar N2 content of the air inside the respirometry chamber remains constant, i.e. it is metabolically inert and does not exchange with N2 inside the organism(s). The water vapour pressure of the air inside the respirometry chamber is determined by the SWVP, which we account for in calculating vg(STPD) (Eqn 3).
To calculate MR, PO2 (from FsO2) must be converted to the fractional O2 content (FO2; %O2/100). In a conventional closed respirometry system (e.g. Vleck, 1987), a dry CO2-free gas sample is analysed at the start (i.e. initial FiO2) and at the end (i.e. final FeO2) of the respirometry period, and the difference is used to calculate V̇O2 (i.e. the PO2 is measured and is converted to FO2 at STPD). The FO2 can also be measured for a second dry sample that still contains CO2 to calculate CO2 production, and it can be measured for a further moist sample to calculate evaporative water loss (Vleck, 1987). In essence, conventional closed-system respirometry is considered an ambient pressure system, with initial and final PO2 values measured at atmospheric pressure and converted to FiO2 and FeO2 (which, as well as vg, are STPD).
However, repeated-measurement fluorometry differs in that the respirometry chamber remains sealed, and the PO2 is measured in situ. This has the advantage of allowing repeated measurements of chamber PO2, but has the disadvantage that any change in Pb within the chamber (e.g. resulting from fluctuating temperature, CO2 accumulation and evaporation or condensation) will proportionally influence the PO2. As far as we can tell, such changing partial pressures between successive measurements are not currently accounted for using commercially available automated fluorescence analysers; indeed, the descriptions regarding the temperature corrections described in the Q2 user manual (Anonymous, 2014) refer to analysis corrections made in response to the fluorometer sensor temperature, not in situ correction.
Changes in water vapour pressure inside the chamber by evaporation or condensation also alter the PO2. In the worst case, an initially dry chamber atmosphere will almost certainly reach H2O vapour saturation throughout a metabolic trial because of evaporation from the organism(s), which would increase internal chamber Pb. This is an important consideration for calculating vg of the chamber prior to sealing it. As such, we advocate allowing the respirometry chamber to reach saturation water vapour pressure prior to sealing, to avoid any effects of subsequent evaporation on barometric pressure in the chamber (hence, measured PO2). The potential errors associated with these approaches are discussed below. Once the respirometry chamber is sealed, however, and assuming the chamber is rigid, changing water vapour pressure affects PO2 and O2 mole fraction in opposite directions. Effectively, changes in PO2 and mole fraction of O2 balance each other out, which is the previously unreported basis of how the Q2 fluorescence analyser measures PO2 independent of chamber humidity (Bello and Bradford, 2016; Bradford et al., 2013; Van Asbrouck and Taridno, 2009).
Accounting for resampling effects
where FsO2,sample denotes the raw standardised value and FsO2,blank denotes the raw standardised value measured for the ‘blank’ chamber.
where FsO2 denotes the raw standardised value, t0 denotes the initial measurement and tn denotes the measurement after n periods.
Repeated sampling complicates the simple calculation of MR by Vleck (1987) using FiO2 and a single FeO2, and we suggest for repeated measurements that the previous O2 measurement be assigned as FiO2 and the next as FeO2 to calculate MR over the elapsed interval between the two measurements. This introduces substantial MR variation as the signal to noise ratio between consecutive measurements of FO2 made over a short period of time is quite low, so we advocate a moving window average (Kenney and Keeping, 1962) or linear regression (Crispin and White, 2013) to account for this variation. This effectively applies a statistical smoothing across the data to further minimise any changes due to noise between consecutive measurements, such as negative MRs, that result from temperature and pressure oscillations in the respirometry chambers, and signal noise. Data averaging may be less necessary for organisms with higher MR.
Biological effects on chamber atmosphere
Following Vleck (1987), RER can be assumed to range between 0.7 and 1; the maximum error associated with assumption of 0.7 or 1.0 is 6.4% and the maximum potential error is 3.2% for an assumed RER of 0.85 (Vleck, 1987).
In summary, there are a number of points where repeated-measures closed-system fluorometric respirometry differs from standard closed-system respirometry. In the latter, CO2 and H2O can be removed prior to measurement of the PO2 (Lighton, 2008; Vleck, 1987), but this is not possible for fluorometric measurements because PO2 is measured inside the sealed respirometry chamber. Rather, any evaporation after sealing the respirometry chamber needs to be accounted for to calculate vg(STPD), changes in PO2 over time by changes in ambient temperature and sensor drift need to be accounted for, and the contribution of CO2 production must be accounted for by assuming an RER. If done so appropriately, in the manner that we have developed here, then the equation for calculation of V̇O2 for repeated-measures fluorometric aerial closed-system respirometry reduces to the same as for conventional closed-system respirometry (Eqn 7).
Measuring RMR in plant seeds
We have used seeds as model organisms to demonstrate the calculation of MR from PO2 measurement by repeated-measures fluorometric closed-system respirometry. We measured seed RMR, which has rarely been done by fluorometric respirometry, but has been done using other methodologies (Dalziell and Tomlinson, 2017; Garwood and Lighton, 1990). However, there are several complications when measuring RMR of seeds that must be considered.
Firstly, metabolic processes in dry seeds are limited by solute mobility, and MR is not readily detectable in seeds with moisture contents in equilibrium with an RH below 90% (Bewley et al., 2013; Walters et al., 2005). Consequently, seeds need to be hydrated sufficiently for metabolic processes to be initiated prior to respiratory measurements being made. Secondly, seeds must not be fully imbibed during measurements of respiration, as germination (a metabolically active process that produces much higher MR compared with quiescent seeds) may occur during the trial. Therefore, in our metabolic trials, to ensure germination would not occur, seeds were pre-equilibrated to 95% RH prior to respirometry measurements. The signal to noise ratio is significantly reduced in quiescent seeds compared with germinating seeds, and to compensate for this we have measured more than one individual seed per chamber. This may offset some of the intrinsic utility suggested for repeated-measures fluorometric respirometry in making single seed measurements (as per Bello and Bradford, 2016), but this is countered somewhat by our ability to measure many replicate chambers at once. Finally, the interactions between seed dormancy status and MR is not well understood (Dalziell and Tomlinson, 2017), and the relationship between seed dormancy and germination/respiration is unclear (Bello and Bradford, 2016; Booth and Sowa, 2001; Bradford et al., 2007; Footitt and Cohn, 1995). As such, we measured RMR of dormant and non-dormant seeds (see methods below).
We measured seeds of five species of Fabaceae native to Western Australia. Using seeds from a single family (i.e. a phylogenetically constrained group) limited the potential influences of embryo type and size, as well as seed dormancy, on seed MR (Dalziell and Tomlinson, 2017). Seeds were collected from wild populations in the Pilbara region of Western Australia between 2010 and 2013 and stored in a controlled environment room at 15°C and 15% RH prior to experiments. Prior to respirometry measurements, all seeds were X-rayed (Faxitron Specimen Radiography System MX-20 Cabinet, Tucson, AZ, USA) to determine seed fill (Erickson et al., 2016). Seeds that did not contain an intact embryo or showed any internal damage were discarded. Seeds were immersed in hot water (95°C) for 2 min to alleviate physical dormancy (Erickson et al., 2016), then blotted dry with paper towel. To reduce the metabolic effect of microbial contamination, all seeds (both dormant and non-dormant) were sealed in porous nylon-mesh bags and soaked in a biocide [2% solution of Plant Preservative Mixture (Plant Cell Technology, Washington, DC, USA), with two drops of the surfactant Tween 20 (Sigma-Aldrich, Castle Hill, NSW, Australia)] for 30 min, after which seeds were rinsed in autoclave-sterilized, reverse-osmosis purified water. All seeds were then placed in respirometry chambers (14 replicate 1800 µl chambers containing seven seeds for each dormant and non-dormant treatment), and placed inside an air-tight electrical enclosure box (NHP Fibox, NHP Electrical Engineering Products, Sydney, Australia) containing a non-saturated solution of LiCl (∼4.8 g per 100 ml H2O; Hay et al., 2008) to allow seeds to equilibrate to 95% RH. The box was placed in an incubator at 20°C for 10 days. To ensure that they had reached 95% RH at the end of the incubation period, non-dormant seeds were tested with a hygrometer (HygroPalm, Rotronic, Crawley, UK). Chambers were then removed from the RH box, capped and measured at 25°C (regulated using an in-built temperature controller) at 30 min intervals over a period of 2 days, using the standard data acquisition software. These data were further analysed according to Eqns 1–7 using the accompanying R script (Script 1). To calculate the initial vg, we used the wet mass of seeds at 95% RH, multiplied by the number of seeds in each respirometry chamber, and Pb was measured using a portable weather station (TechBrands XC0394, Electus Distributions, Rydalmere, NSW, Australia). Subsequent to measurement, to ensure successful dormancy break and seed viability, all seeds were removed from the chambers and plated in Petri dishes containing water agar (0.7% w/v), placed in an incubator at 25°C, and assessed for germination over a period of 28 days.
The majority of seed RMR should result from the combined activity of the cotyledons, plumule and radicle, which we broadly considered as the ‘embryo’. To compare the RMR of seeds between species, and to understand the interspecific effects of seed mass and embryo mass, dry masses of whole seeds and excised embryos were determined. To determine embryo masses, a subsample of 30 non-dormant seeds of each species were imbibed for 24 h, and the embryos were excised using a pair of fine forceps and a scalpel under a binocular microscope. The embryos and the seed coats were separated, and split into three replicates (i.e. 10 seeds per replicate) and placed in individual metal pans and dried for 48 h at 105°C. Average dry masses of the 30 embryos and remaining seed contents were measured using a five-point balance (Mettler Toledo, Switzerland).
RESULTS AND DISCUSSION
RMR of angiosperm seeds
Oxygen depletion was linear for all species and resulted in constant RMR following our calculations. RMR ranged between 0.26±0.062 and 2.01±0.71 µl O2 h−1 for non-dormant seeds. Dormant seeds had RMR approximately 53% lower than this (Table 1). These values are within the range reported for aerobic metabolism of seeds, showing similar metabolic depression in dormant seeds (Dalziell and Tomlinson, 2017; Garwood and Lighton, 1990).
Biological variation in RMR
The biological sources of error described below contribute to both interspecific and intraspecific variation in MR. Although these form the basis of the biological patterns that most ecologists and evolutionary biologists are interested in, they also contribute potential errors to the measurement of MR, and need to be appropriately addressed (Felsenstein, 1985).
Mass is the most pervasive influence on MR of organisms, overpowering the combined influences of ecology and evolution (Darveau et al., 2002; Gillooly et al., 2001). Previous repeated-measures fluorometric respirometry studies (Bello and Bradford, 2016; Bradford et al., 2013) have not considered the influence of seed mass, which has substantial implications for the calculation of initial gas volume and the expression of MR corrected for metabolically active mass. Calculation of RMR in absolute terms (i.e. millilitres O2 consumed per hour) does not account for the potentially complex allometric effects of mass that have a pervasive influence on RMR (Dalziell and Tomlinson, 2017). RMR of seeds (Table 1) shows that, even in closely related taxa, mass correction and embryo mass correction markedly influence the species comparisons. We have avoided phylogenetic influence to a large degree by measuring closely related species, but phylogenetically independent contrast analyses (Felsenstein, 1985) should be considered for broader interspecific comparisons, especially because the phylogenetic diversity of plant seeds is greater than that of any animal group studied to date. Finally, we specifically avoided using seeds with complex dormancy types, as there is a limited understanding about how these kinds of dormancy may affect RMR, and populations of seeds with physical dormancy are simple to manipulate into non-dormant fractions. Nevertheless, differences in RMR between dormant and non-dormant seeds (Table 1) are interspecifically variable, even in closely related species, and patterns are variable depending on different mass corrections. Investigating these patterns was not our aim here, but the data that we have presented indicate that a substantial amount of research is required to do so.
Methodological errors can contribute biological errors, e.g. some seeds became contaminated with fungal and/or bacterial infections. Alternatively, when dealing with multiple individuals in a single chamber, some replicates might include more-or-less metabolically inactive individuals. These errors result in much quicker or slower O2 depletion rates, respectively. We have removed trials where O2 depletion rates were greater than two standardised residuals different to average O2 depletion rates (Dataset 1). We have not specifically included this process in the calculation algorithm (Script 1) because this is one of several statistical approaches to identify outliers.
Respirometry errors
As previously stated, although the biological variation described above contributes to patterns of interest to ecologists and evolutionary biologists, these patterns need to be appropriately accounted for so as to not influence the MR measured by repeated-measures fluorometric aerial closed-system respirometry.
Organism density and chamber gas volume
To account for organism volume in the respirometer (Vleck, 1987), mass is corrected to volume by using an estimate of organism density. Seed density typically ranges between 1.0 and 1.5 g cm−3 (Gupta and Das, 1997; Nelson, 2002). Resulting errors in gas volume by not correcting organismal volume can be considerable; for four 100 mg seeds in a 2200 µl respirometry chamber, error ranges from an ∼16% (ρ=1.0) to 24% (ρ=1.5) overestimate of V̇O2 (Fig. 1A). As such, we advocate an appropriate organism mass density be accounted for and suggest assuming (if necessary) the median of published values of 1.2 g cm−3 for seeds. This results in a nominal error of approximately 1.25% if ρ is between 1.0 and 1.5 g cm−3.
Water vapour pressure
The next potential error in the calculation of V̇O2 also relates to vg, on the basis of initial water vapour density, to avoid the influences of initial and changing water vapour pressure on vg. Because saturation water vapour pressure is temperature dependent (Parrish and Putnam, 1977), the water present in saturated air changes with temperature, and hence error is also temperature dependent. Exploring the worst-case scenario where chamber atmosphere was initially dry (0% RH) but we assume saturation, the vg error at 10°C is approximately 1% underestimated (Fig. 1B). As temperature increases to 40°C, this increases to approximately 7%. In a more realistic scenario, where chamber humidity is actually 50%, the error is approximately 2.7% at 10°C, and increases to 13% at 40°C. Although these errors are small at biologically appropriate temperatures, they need to be minimised to standardise MR measurements, especially under experimental treatments where a broader range of temperatures may be applied.
Respiratory exchange ratio and O2 limitation
We have made several assumptions relating to PCO2, and how this influences the changing conditions in the respirometry chamber and the subsequent effects on quantifying MR. The first effect, that initial FCO2 is zero, is minor if the initial FCO2 is actually at atmospheric levels, resulting in <0.04% error in vg (current atmospheric FiCO2 is 0.000406; https://www.esrl.noaa.gov/gmd/ccgg/trends/). The second effect, where CO2 accumulates throughout the measurement period, is substantial, if accumulated metabolic CO2 production is not accounted for. RER is usually between 0.7 and 1.0, and the maximum possible errors in calculated V̇O2 will occur when estimated RER=0.7 when actually RER=1, or vice versa (Fig. 1C) (Vleck, 1987). Overestimates of V̇O2 result if actual RER is less than assumed RER, and underestimates result if the reverse is true, with a maximum error of approximately 6.5% (Vleck, 1987). On this basis, minimal error is obtained by assuming RER=0.85, restricting the maximum possible error to ±3.2% (Vleck, 1987). Prolonged closure of the metabolic system can cause more substantial errors of approximately 10–13% because of the greater storage capacity of organismal water for CO2 than for O2 (Malte et al., 2016).
Hypoxia has a potential effect on MR per se. A rule of thumb for terrestrial animals suggests that O2 levels below 12% (atmospheric) begin to affect MR (Bradford and Seymour, 1988; Seymour et al., 1995; Wheatly, 1981); 12% O2 is equivalent to approximately 58% of initial O2 using the standardised Q2 scale. As O2 becomes limiting in the chamber, it decreases MR over time (Fig. 2A), resulting in an asymptotic O2 depletion curve rather than the expected straight line. This can be corrected by analysing data prior to the time at which hypoxia may have occurred. However, during dynamic state applications, such as germination, O2 limitation generates unpredictable outcomes. Peak MR, such as that estimated at the midpoint of O2 depletion curves during germination (R50; Bello and Bradford, 2016; Bradford et al., 2013) are potentially underestimated because they have been measured in hypoxic atmospheres and this potentially reduces the MR of the seeds.
Conclusions and broader applicability
The approaches to utilising repeated-measures fluorometric aerial closed-system respirometry explored here potentially expand the comparative approach to seed biology, enabling quantitative evaluation of metabolic adaptations to a range of environmental and biological factors, as explored in animals (Bartholomew, 1964; Cloudsley-Thompson, 1991; Degen, 1997). Although evolutionary patterns in the MR of animals are well documented and continue to be debated (Uyeda et al., 2017), they have yet to be explored for plant seeds. Although our study reveals the complexity of calculating accurate RMR from repeated-measures fluorometric respirometry, even greater and less predictable complexity is entailed in making dynamic state calculations, and these will require further exploration. Nevertheless, repeated-measurement fluorescence sensors extend the capabilities of closed-system respirometry in ways that potentially revolutionise the study of MR for small taxa. Although measurement of V̇CO2 using infra-red gas analysers has been accomplished for quite small taxa (Ayton et al., 2016; Lighton and Fielden, 1995; Terblanche et al., 2004), fluorometric closed-system respirometry has the capacity to extend this lower size limit substantially, including different life history stages such as eggs and larvae (Köster et al., 2008; Szela and Marsh, 2005). Measurements of the MR of such organisms, particularly as a response of thermal performance, are often crucial to understanding ecophysiological responses to changing environments (Kearney et al., 2010; Tomlinson et al., 2018). As energetics is the basis of many processes structuring ecosystems (McNab, 2002; Tomlinson et al., 2014), we expect great potential insights to be gained from measurement of MR using repeated-measurement fluorometric respirometers for small organisms.
Acknowledgements
Dr Shane Turner provided useful insight into seed physiology throughout the development of this analysis.
Footnotes
Author contributions
Conceptualization: S.T., E.L.D., D.J.M.; Methodology: S.T., E.L.D., P.C.W., W.L., D.J.M.; Software: S.T., W.L.; Investigation: S.T., E.L.D., P.C.W., W.L.; Resources: D.J.M.; Writing - original draft: S.T.; Writing - review & editing: S.T., E.L.D., P.C.W., W.L., K.W.D., D.J.M.; Funding acquisition: S.T., E.L.D., P.C.W., K.W.D., D.J.M.
Funding
This research is an outcome of Australian Research Council grant LP160100381, which provided financial support for E.L.D. Financial support for the Q2 was provided by the BHP Billiton Iron Ore Community Development Project (contract no. 8600048550) under the auspices of the Restoration Seed Bank Initiative, a partnership between BHP Billiton Iron Ore, The University of Western Australia, and the Botanic Gardens and Parks Authority. S.T. was supported by an Australian Government unemployment allowance and latterly by the Australian Research Council grant IC150100041.
References
Competing interests
The authors declare no competing or financial interests.