ABSTRACT
Despite progress made in confocal microscopy, even fast systems still have insufficient temporal resolution for detailed live-cell volume imaging, such as tracking rapid movement of membrane vesicles in three-dimensional space. Depending on the shortfall, this may result in undersampling and/or motion artifacts that ultimately limit the quality of the imaging data. By sacrificing detailed information in the Z-direction, we propose a new imaging modality that involves capturing fast ‘projections’ from the field of depth and shortens imaging time by approximately an order of magnitude as compared to standard volumetric confocal imaging. With faster imaging, radiation exposure to the sample is reduced, resulting in less fluorophore photobleaching and potential photodamage. The implementation minimally requires two synchronized control signals that drive a piezo stage and trigger the camera exposure. The device generating the signals has been tested on spinning disk confocal and instant structured-illumination-microscopy (iSIM) microscopes. Our calibration images show that the approach provides highly repeatable and stable imaging conditions that enable photometric measurements of the acquired data, in both standard live imaging and super-resolution modes.
This article has an associated First Person interview with the first author of the paper.
Footnotes
Author contributions
Conceptualization: X.H., O.B.; Methodology: X.H.; Software: F.M.; Validation: F.M.; Formal analysis: F.M.; Investigation: X.H., S.J., F.M.; Resources: X.H., M.P.S., O.B., F.M.; Data curation: X.H.; Writing - original draft: X.H., F.M.; Writing - review & editing: X.H., S.J., M.P.S., O.B., F.M.; Visualization: X.H., S.J., F.M.; Supervision: X.H., M.P.S., O.B.; Project administration: X.H., O.B.; Funding acquisition: O.B., F.M.
Funding
This work was supported by funding from the National Research Foundation Singapore, MechnoBiology Institute of Singapore (R-714-105-004-135) as well as The Research Council of Norway (RCN; Norges Forskningsråd) (RCN INFRASTRUKTUR 208310/F50).